Skip Navigation



NDT Advance Access published online on February 1, 2005

Nephrology Dialysis Transplantation, doi:10.1093/ndt/gfh676
This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
20/4/754    most recent
gfh676v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Teraoka, T.
Right arrow Articles by Ueda, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Teraoka, T.
Right arrow Articles by Ueda, S.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Nephrol Dial Transplant © ERA-EDTA 2005; all rights reserved
Received September 6, 2004
Accepted November 19, 2004


Original Articles

Prediction of iopromide reduction rates during haemodialysis using an in vitro dialysis system

Tomohisa Teraoka 1*, Takao Sugai 2, Sunao Nakamura 2, Hiroyuki Hirasawa 3, Shigeto Oda 3, Hidetoshi Shiga 3, Mio Suga 1, Shinji Yamane 4, Hiroyuki Ishii 4, Shin-ichi Yamagata 1, Nobunori Satoh 1, and Shiro Ueda 1

1 Department of Drug Information and Communication, Graduate School of Pharmaceutical Sciences, Chiba University, Chiba, Japan
2 Shin-Tokyo Hospital, Chiba, Japan
3 Department of Emergency and Critical Care Medicine, Chiba University School of Medicine, Chiba, Japan
4 Division of Blood Purification, Chiba University Hospital, Chiba, Japan

* To whom correspondence should be addressed.
Tomohisa Teraoka, E-mail: tera{at}p.chiba-u.ac.jp



  Abstract

Background. In clinical studies, it has been difficult to evaluate the influence of haemodialysis (HD) parameters on HD clearance (CLHD) and reduction rate (RR) of non-ionic contrast medium during HD sessions. We therefore predicted clinical values of CLHD and RR of iopromide, a non-ionic contrast medium, from findings obtained from in vitro experiments, and confirmed that these predictive values were comparable with the actual values in clinical cases.

Methods. We developed a correlation equation for predicting CLHD on the basis of in vitro HD experiments by varying blood flow rates between 100 and 200 ml/min with a cuprammonium rayon dialyser (AM-SD-10H). Total body clearance of iopromide (CLPT) was estimated by the Cockroft-Gault equation. The volume of distribution (Vd) was obtained from the reported value. By using the HD and three pharmacokinetic parameters (CLHD, CLPT and Vd), we predicted CLHD and RR for seven patients undergoing HD after the administration of iopromide.

Results. In the in vitro study, the mean values (±SD) of iopromide clearance at blood flow rates of 100, 150 and 200 ml/min were 45.35 (2.54), 53.88 (6.46) and 57.61 (4.72) ml/min, respectively. There were highly significant correlations between clearance and blood flow rate (r = 0.975). Although the predicted CLHD showed a tendency towards underestimation, a good correlation was found. Predicted RR values were similar to observed values except for one case.

Conclusion. The in vitro model used in the present study provides pertinent information about CLHD and is helpful for predicting RR during HD in individual patients undergoing HD.

Keywords: drug removal; haemodialysis; iopromide; pharmacokinetics; radiocontrast media.
Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.