Skip Navigation


NDT Advance Access originally published online on December 15, 2008
Nephrology Dialysis Transplantation 2009 24(5):1550-1553; doi:10.1093/ndt/gfn696
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
24/5/1550    most recent
gfn696v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Markota, I.
Right arrow Articles by Tomic, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Markota, I.
Right arrow Articles by Tomic, M.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author [2008]. Published by Oxford University Press on behalf of ERA-EDTA. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org



Measuring of the heparin leakage into the circulation from central venous catheters—an in vivo study

Ivica Markota1, Darko Markota2 and Monika Tomic3

1 Division of Nephrology and Dialysis, Department of Medicine, Mostar University Hospital Center 2 Department of Medicine, Mostar University Hospital Center 3 Department of Internal Medicine, Mostar University Hospital Center, University of Medicine Mostar, Mostar, Bosnia and Herzegovina

Correspondence and offprint requests to: Ivica Markota, Department of Medicine, Mostar University Hospital Center, 88 000 Mostar, Bosnia and Herzegovina. Tel: +387-36-341-955; Fax: +387-36-341-974; E-mail: markota.ivica{at}tel.net.ba, ivica_markota{at}net.hr



  Abstract

Background. A catheter lock with a highly concentrated heparin solution is often used to maintain its patency. The result of the in vitro study shows a significant catheter leakage that occurs after locking the catheter. The goal of this study is to develop a model to measure the catheter leakage in vivo and test it on various kinds of catheters.

Methods. Twenty-four patients with central venous dialysis catheters were examined. After the 48-h interdialytic period, we aspirated the contents of the catheter lumen for analysis. We simultaneously took a sample of the peripheral blood for analysis. In the second part of the test, instead of taking the sample after 48 h, we took it after 10 min. Based on the difference in haematocrit in those two samples, we were able to determine the amount of heparin that remained in the catheter, and indirectly, the amount of heparin that leaked out of the catheter.

Results. Using the lock volumes indicated on the catheter by the manufacturer, the early leakage is significantly higher in nontunnelled catheters compared to tunnelled Hemoflow and Ash Split catheters (P = 0.05). There is no significant statistical difference in the early leakage between Ash Split and Hemoflow catheters. The late leakage is significantly higher in nontunnelled catheters compared to Hemoflow and Ash Split catheters (P = 0.05). There is no significant statistical difference in the total leakage between Ash Split and Hemoflow catheters.

Conclusion. We present a model that enables the measurement of the catheter leakage in vivo. We applied the model on three kinds of catheters and concluded that both early and late leakages are significantly higher in nontunnelled catheters compared to Hemoflow and Ash Split tunnelled catheters. Our results show that the so-called early leakage measured in vivo is significantly lower compared to the results from in vitro studies. Further research is necessary to determine the amount of leakage volume for different kinds of catheters and to determine the exact leakage dynamics of lock solutions in vivo.

Keywords: catheter lock; haemodialysis; heparin leakage; heparin

Received for publication: 10. 7.08
Accepted in revised form: 21.11.08


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Nephrol Dial TransplantHome page
H.-D. Polaschegg
Measuring lock solution spillage in-vivo versus in-vitro
Nephrol. Dial. Transplant., July 1, 2009; 24(7): 2294 - 2295.
[Full Text] [PDF]


Home page
Nephrol Dial TransplantHome page
I. Markota
Reply
Nephrol. Dial. Transplant., July 1, 2009; 24(7): 2295 - 2295.
[Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.