Skip Navigation

This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (7)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Wiese, P.
Right arrow Articles by Nonnast-Daniel, a. B.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Wiese, P.
Right arrow Articles by Nonnast-Daniel, a. B.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Nephrol Dial Transplant (2003) 18: 1397-1400
© 2003 European Renal Association-European Dialysis and Transplant Association


Technical Note

Clinical and Doppler ultrasonography data of a polyurethane vascular access graft for haemodialysis: a prospective study

Patrick Wiese1, Jochen Blume1, Hans-Joachim Mueller1, Hermann Renner2 and and Barbara Nonnast-Daniel1,

1 Department of Nephrology and 2 Department of Vascular Surgery, Klinikum Nuernberg, Nuernberg, Germany

Abstract

Background. Absence of a permanent vascular access in most patients starting haemodialysis remains a cause of high morbidity and costs. This study obtained new clinical and colour Doppler ultrasound (CDU) data of a polyurethane vascular access graft (PVAG) proposing early post-operative cannulation.

Methods. Baseline characteristics were determined in 15 patients and the PVAGs were evaluated prospectively including first cannulation, patency and complications. CDU was used post-operatively and after 1 year for assessing graft morphology and access blood flow.

Results. PVAGs were cannulated at a median of 4 days post-operatively. The 1-year primary patency of the PVAG was 66.7%. During the 15 months observation three grafts thrombosed, one was replaced because of infection and one because of ischaemia. CDU measurements at the feeding brachial artery revealed a mean initial access volume flow of 773±89 ml/min, being significantly higher in patients without thrombosis compared to patients with thrombotic events (930±90 vs 375±143 ml/min, P<0.05). The initial inability to directly monitor PVAGs by CDU changed at sites of frequent centesis, where Doppler signals and luminal morphology could be evaluated in the follow up examination.

Conclusions. The PVAG offers early access for urgent haemodialysis. CDU for access volume flow measurement at the feeding brachial artery contributes to predict access thrombosis. Direct non-invasive graft imaging is limited and the ultrasonographical changes in the polyurethane material enabling graft monitoring after repeated cannulation might indicate an injury of the graft with increased risk for access failure.

Keywords: Doppler ultrasonography; early cannulation; haemodialysis; polyurethane graft; prospective study; vascular access

Notes

Correspondence and offprint requests to: Barbara Nonnast-Daniel, MD, University Erlangen-Nuernberg, Medizinische Klinik IV/4, Klinikum Nuernberg, Breslauer Strasse 201, D-90471 Nuernberg, Germany. Email: werner.g.daniel{at}t-online.de


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Nephrol Dial TransplantHome page
P. Wiese and B. Nonnast-Daniel
Colour Doppler ultrasound in dialysis access
Nephrol. Dial. Transplant., August 1, 2004; 19(8): 1956 - 1963.
[Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.