Skip Navigation

This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (3)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Külz, M.
Right arrow Articles by Schneider, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Külz, M.
Right arrow Articles by Schneider, H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Nephrol Dial Transplant (2002) 17: 1475-1479
© 2002 European Renal Association-European Dialysis and Transplant Association

In vitro and in vivo evaluation of a new dialyzer

Michael Külz1, Bernd Nederlof1 and Hans Schneider2,

1 Fresenius Medical Care, Bad Homburg, Germany and 2 Nephrology Center Stuttgart, Stuttgart, Germany

Background. Recent dialyzer-related developments have concentrated upon the improvement of performance and biocompatibility. The focus of these developments was predominantly on the membrane itself. A newly developed high-flux dialyzer (FX60) with an advanced Fresenius Polysulfone® membrane (Helixone®) overcomes this limitation and has several design-related advantages attributed to the redesign of the individual functional components. For the first time, polypropylene was selected as the material for the dialyzer housing. Both the fibre and the fibre-bundle geometry were refined with the aim of improving overall performance and to reduce dialysate consumption.

Methods. This study aims at investigating the in vitro and in vivo performance of the new FX60 dialyzer with the focus on dialysate flow distribution and dialysate consumption. A new method to analyse dialysate flow distribution, based on local clearances, is suggested. The effect of reducing dialysate flow from 500 to 300 ml/min is investigated in vivo.

Results. K0Aurea, a common measure to quantify device performance, is found to approach 1000 ml/min for the FX60 with a surface of only 1.4 m2. Local clearance measurement shows equal performance of the Helixone® fibre bundle over the entire cross-section. A dialysate flow reduction by 20% (in vivo) only results in a minor loss in clearance.

Conclusions. The newly developed high-flux dialyzer (FX60) shows a remarkable performance (K0Aurea). The excellent utilization of dialysate could be proven in vivo and is attributed to the superior dialysate flow distribution. A reduction of dialysate flow by 20% could lead to substantial economic savings.

Keywords: dialysate consumption; dialysate flow distribution; high-flux dialyzer; local clearance; polysulfone membrane

Correspondence and offprint requests to: Prof. Dr H. Schneider, Nephrology Center Stuttgart, Wolframstraße 60, D-70191 Stuttgart, Germany. Email: tabeti{at}t\|[hyphen]\|online.de


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.