NDT Advance Access originally published online on May 23, 2006
Nephrology Dialysis Transplantation 2006 21(9):2439-2445; doi:10.1093/ndt/gfl249
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
© The Author [2006]. Published by Oxford University Press on behalf of ERA-EDTA. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org
Original Articles: Clinical Nephrology
Does the ID-MS traceable MDRD equation work and is it suitable for use with compensated Jaffé and enzymatic creatinine assays?
1 Department of Clinical Biochemistry, 2 Department of Renal Medicine, East Kent Hospitals NHS Trust, Kent and Canterbury Hospital, Canterbury, Kent, UK, 3 The WellChild Laboratory, King's College London, Department of Paediatrics, Guy's Hospital, London, UK and 4 The AASK Core Biochemistry Laboratory, Department of Clinical Pathology, Cleveland Clinic Foundation, Cleveland, Ohio, USA
Correspondence and offprint requests to: Dr Susan Vickery, Department of Clinical Biochemistry, East Kent Hospitals NHS Trust, Kent and Canterbury Hospital, Canterbury, Kent CT1 3NG, UK. Email: susan.vickery{at}ekht.nhs.uk
| Abstract |
|---|
|
|
|---|
Background. International recommendations suggest that measurement of serum creatinine should be supplemented with an estimate of glomerular filtration rate (GFR) using the Modification of Diet in Renal Disease (MDRD) study equation. One problem has been the lack of standardization of commercially available creatinine assays resulting in varying estimates of GFR. A revision of the MDRD equation offers traceability to a reference method. This study evaluates the use of isotope dilution mass spectrometry (ID-MS), the compensated Jaffe and enzymatic creatinine methods compared with the Beckman CX3 Jaffe assay used to derive the MDRD equation and investigates their impact on GFR estimation using both the original and ID-MS-traceable MDRD equations.
Methods. Serum creatinine was measured in 277 patients by (i) ID-MS, (ii) a Roche enzymatic assay, (iii) a Roche compensated kinetic Jaffe assay and (iv) a Beckman CX3 kinetic Jaffe assay. Estimated GFR was calculated using the MDRD equations.
Results. The ID-MS (7.5%), Roche enzymatic (8.6%) and compensated kinetic Jaffe (11.9%) assays were all negatively biased (P < 0.0001) compared with the Beckman CX3 assay, causing predictable, clinically significant, overestimation of GFR when the original MDRD equation is used. This positive bias was reduced (ID-MS 6.7 to 0.4%; enzymatic 8.8 to 3.4%; compensated kinetic Jaffe 13.7 to 7.1%) when GFRs were calculated using the ID-MS-traceable MDRD equation.
Conclusions. Compensated assays that account for non-creatinine chromogen interference produce significantly higher estimates of GFR when using the original MDRD equation. Use of the ID-MS-traceable MDRD equation ameliorates this effect. There is good agreement between estimated GFR derived from the original MDRD equation using Beckman Astra CX3 data and estimated GFR derived from the new ID-MS-traceable MDRD equation using a local ID-MS creatinine assay. This suggests that the ID-MS-traceable MDRD equation may be reliably used with both ID-MS and true ID-MS-traceable creatinine assays without the requirement for standardization to the MDRD laboratory.
Keywords: chronic kidney disease; creatinine; glomerular filtration rate; isotope dilution mass spectrometry; Jaffe assay; MDRD equation
| Introduction |
|---|
|
|
|---|
International recommendations suggest that measurement of serum creatinine should be supplemented with an estimate of the glomerular filtration rate (GFR) and the Modification of Diet in Renal Disease (MDRD) study equation [1,2] is generally thought to be the least biased and most accurate method of achieving this [3,4]. It is well understood that the equation is susceptible to changes in the standardization of creatinine assays, particularly at creatinine concentrations in the normal to high-normal range [58]. The development of the MDRD equation was undertaken using a kinetic Jaffe assay on a Beckman Astra CX3 analyser [1]. Recently, a revised isotope dilution mass spectrometry (ID-MS) traceable version of the equation was published, validated for use with methods demonstrating zero or minimal bias compared with ID-MS [9].
In addition to known problems of calibration [1012], creatinine assays that use the Jaffe reaction are also known to be susceptible to interference from non-creatinine (pseudo) chromogens, which may account for up to 20% of the apparent creatinine component of normal serum. Enzymatic creatinine assays do not typically suffer from this interference. In an attempt to improve this situation, a major laboratory reagent manufacturer, Roche Diagnostics Ltd, have introduced the so-called compensated Jaffe assay for measurement of serum creatinine, in which a fixed component is automatically subtracted from each result to account for reaction due to non-creatinine chromogens.
In the present study, we have compared creatinine results produced by both the Roche compensated Jaffe and enzymatic assays with those produced by the Beckman Astra CX3 assay in the MDRD laboratory, and have evaluated their practical application in estimating GFR using both the original and ID-MS-traceable MDRD equations. Verification of true creatinine concentration was achieved using an ID-MS approach, which has been proposed as the reference method [13].
| Subjects and methods |
|---|
|
|
|---|
Serum samples were available from 277 patients, including 203 patients enrolled in a study of chronic kidney disease (CKD), the clinical details of whom have been reported elsewhere [14], and 74 residual, de-identified samples from the diagnostic laboratory to provide a representative range of creatinine concentrations.
Serum creatinine concentrations were measured by the following:
- ID-MS, a modified liquid chromatographic ID_MS/MS method [15]. Serum (5 µl) was diluted with 250 µl of deionized water containing 500 pmol d3-creatinine and 250 µl acetonitrile with 0.05% formic acid. Following mixing and centrifuging, supernatant (2 µl) was automatically injected using an HTS PAL autosampler (CTC Analytics AG, Switzerland) into a 200 µl/min solvent stream of acetonitrile:water (50:50 v/v) with 0.025% (v/v) formic acid. Chromatography was performed on a Symmetry-C8, 3.5 µm, 2.1 x 50 mm column (Waters Corporation, UK) and precursorproduct ion pairs (m/z 114.2/44.2, 117.2/47.2) were acquired in positive-ion multiple-reaction-monitoring mode using an API4000 (Applied Biosystems, UK). Results were calculated using Analyst version 1.3.1. Stock aqueous creatinine standard, 10 mmol/l, was prepared by dissolving creatinine [BDH (Chemicals) Ltd, UK] in 0.1 M HCl and stored at 4°C. The concentration of the stock standard has been confirmed using National Institute of Standards and Technology (NIST) standard reference material, creatinine 914a (Laboratory of the Government Chemist, UK). Assay calibrators at 50, 250 and 1000 µmol/l were prepared from the stock standard by dilution with deionized water. Between-day coefficients of variation in control sera were <5% at concentrations of 112 and 256 µmol/l.
- A kinetic Jaffe assay on a Beckman Astra CX3 analyser (Beckman-Coulter Ltd, UK). These analyses were provided by the African-American Study of Kidney Disease (AASK) Core Biochemistry laboratory (Cleveland Clinic Foundation), which supported the MDRD study [1]. Between-day coefficients of variation were 10.0 and 3.7% at concentrations of 95 and 360 µmol/l. The method uses two aqueous calibrators at values of 88 and 707 µmol/l. Absorbances are read at 520 and 560 nm, 25.6 s after sample addition. The assay has been reported to have an average 6 µmol/l positive bias compared with a high-performance liquid chromatography (HPLC) procedure [16]. For this assay only, samples were analysed in duplicate and data were analysed using the mean of duplicates.
- A kinetic compensated Jaffe assay on an Integra 800 analyser (cat. no. 20764345, Roche Diagnostics Ltd, UK). This is a rate assay without deproteinization, measuring the increase in absorbance at 512 nm between 55 and 70 s after initiation of the reaction. Absorbance blanking at 583 nm is used. The assay is calibrated with a lyophilized human serum calibrator (C.f.a.s., Roche Diagnostics Ltd) in which the creatinine concentration (normally
320 µmol/l) is traceable to an ID-MS determination. To compensate for non-creatinine chromogens, values are automatically corrected by subtracting 18 µmol/l. The between-day coefficients of variation were 4.4 and 3.9% at concentrations of 93 and 316 µmol/l, respectively.
- An enzymatic wet chemistry method on an Integra 800 analyser (creatinine plus ver. 2, cat. no. 03263991, Roche Diagnostics Ltd). The assay principle utilizes a creatininasecreatinasesarcosine oxidase system with detection at 552 nm and absorbance blanking at 659 nm. The assay is calibrated with C.f.a.s. (Roche Diagnostics Ltd) as above. Water is used as the zero standard. The between-day coefficients of variation were <2.5% at concentrations of 109 and 371 µmol/l.
Estimated GFR was calculated using the original MDRD equation [2]: GFR (ml/min/1.73 m2) = 186x[creatinine (µmol/l) x 0.011312]1.154 x (age)0.203 x 0.742 (if female) x 1.212 (if black) in addition to the ID-MS-traceable MDRD equation [9] in which the constant 186 is replaced by the constant 175.
Statistical analysis
All analyses were undertaken using Analyse-ItTM (Analyse-It Software Ltd, Leeds, UK) and InStat (GraphPad Software Inc., San Diego, CA, USA). To determine whether it is appropriate to use the Roche Jaffe or enzymatic assays in the MDRD equation, and to check the validity of the ID-MS-traceable equation using a local ID-MS creatinine assay, the Beckman Astra CX3 results were used as the reference method against which other methods were compared by studying bias (plots of difference between test method and Beckman method plotted against Beckman results) and agreement (linear regression analysis). The significance of bias compared with the Beckman method was tested using the Wilcoxon's matched-pairs signed-ranks test, and the relationship between bias and concentration was tested using linear regression analysis. Bias of the three colorimetric methods against ID-MS was also tested. Each of the serum creatinine data sets had a non-parametric distribution, as confirmed using the ShapiroWilk W test. Therefore, log-transformed data were used for bias analyses [17]. To illustrate the effect of assay bias on estimated GFR, the regression equations were used to generate creatinine results for all three assays relative to the Beckman assay at 10 µmol/l intervals covering the range 80250 µmol/l. These data were then used to generate illustrative estimates of GFR using both the original and ID-MS-traceable MDRD equations assuming the subject to be a 60-year-old Caucasian male. The significance of differences in estimated GFRs (eGFRs) was tested using Student's t-test. Amongst the subset of 203 chronic kidney disease (CKD) patients, where age, gender and ethnicity were known, eGFRs were calculated using the original and ID-MS-traceable MDRD equations for each creatinine method; data were compared using bias plot analyses.
| Results |
|---|
|
|
|---|
Comparison of creatinine methods
The ID-MS (7.5%), Roche enzymatic (8.6%) and compensated kinetic Jaffe (11.9%) methods all produced creatinine results that were negatively biased (P < 0.0001) compared with those obtained from the Beckman Astra CX3 assay (Table 1, Figure 1). The relatively similar negative intercepts of the ID-MS (16.3 µmol/l), Roche enzymatic (16.5 µmol/l) and compensated kinetic Jaffe (17.5 µmol/l) assays compared with the Beckman Astra CX3 assay suggest that a large component of this bias may relate to pseudo-chromogen interference in the Beckman assay, which does not affect the ID-MS or enzymatic assays and has been compensated for in the Roche Jaffe assay. In keeping with this suggestion, there was a constant relationship between bias (testBeckman Astra CX3) and creatinine concentration for the ID-MS method. For both Roche assays the difference increased with increasing creatinine concentration, suggesting an additional component of calibration bias (Figure 1). In the case of the enzymatic assay, a weak, but significant, concentration-bias effect (R2 = 0.06, P < 0.0001; difference = 0.0173 x creatinine concentration 16.5) was seen. A greater concentration-bias effect was seen with the compensated Jaffe assay (R2 = 0.41, P < 0.0001; difference = 0.0434 x creatinine concentration 17.2).
|
|
The Roche compensated kinetic Jaffe assay demonstrated significant negative bias against ID-MS [mean bias = 4.7%, 95% confidence interval (CI) 5.5 to 4.0%, P < 0.0001]. Although the Roche enzymatic assay demonstrated statistically significant negative bias compared with ID-MS, in practice the differences were minimal (mean bias = 1.3%, 95% CI 2.10.0%, P < 0.0001).
Estimated GFRs using an illustrative dataset
Estimated GFRs derived from the ID-MS, Roche enzymatic and compensated kinetic Jaffe creatinine assays were significantly greater than those derived using the Beckman CX3 assay, irrespective of whether the original or ID-MS-traceable MDRD equation was used (P < 0.005 in all cases) (Table 2). The greatest percentage differences were seen at lower creatinine concentrations: as creatinine concentration increases the percentage overestimation of GFR decreases for all three methods compared with the Beckman Astra CX3 assay. Using the original MDRD equation, all three assays produced GFR estimates with >15% positive bias compared with the Beckman assay in the creatinine range 80150 µmol/l. The overestimation was ameliorated, though not abolished, when the ID-MS-traceable MDRD equation was used (Table 2).
|
Effect of ID-MS-traceable MDRD equation on estimated GFR in CKD patients
The ID-MS, Roche enzymatic and compensated kinetic Jaffe assays all produced estimated GFRs, calculated by the original MDRD equation, that were positively biased compared with eGFRs produced by the Beckman Astra CX3 assay (Table 3). This positive bias was reduced (ID-MS 6.7 to 0.4%; enzymatic 8.8 to 3.4%; compensated kinetic Jaffe 13.7 to 7.1%) when GFRs were calculated using the ID-MS-traceable MDRD equation, and there was very good agreement between GFRs calculated from our local ID-MS method using the new MDRD equation and GFRs calculated from the MDRD laboratory method using the original MDRD equation.
|
| Discussion |
|---|
|
|
|---|
Increasing use of eGFR has re-focused the attention of the scientific community on the shortcomings of creatinine methodology. In line with the National Kidney Diseases Education Programme (NKDEP), the Australasian Creatinine Consensus Working Group [18] recommend that serum creatinine assays should be considered acceptable with respect to bias if their results lie within ±15% of values calculated by ID-MS. Over a broad range of creatinine concentrations the Beckman CX3 assay fulfils this accuracy criterion, as do the majority of commercial kinetic Jaffe assays. At lower concentrations few assays do, largely due to the pseudo-chromogen effect [12]. The NKDEP has set a goal of an overall accuracy of ±30% for the estimation of GFR so as to encompass components of analytical inaccuracy and inaccuracies inherent in the equation. Although there was no gold standard estimate of GFR in the present study, it is clear that analytical differences alone can generate GFR estimates in excess of this goal at lower creatinine concentrations; for the assays we studied, the problem was ameliorated by using the more appropriate, ID-MS-traceable MDRD equation.
We have used a direct approach to estimate the errors in estimated GFR associated with use of a compensated kinetic Jaffe assay, when compared with the kinetic Jaffe assay originally used to derive the MDRD equation. Although there are significant concerns relating to the precision of GFR estimates, we concentrated, in this study, on issues of relative accuracy. This is currently compromised by the lack of standardization between creatinine assays and non-consistent correction for pseudo-chromogens. The main clinical consequence of this will be differential recognition of CKD stage 3 (GFR 3059 ml/min/1.73 m2), where classification is based upon GFR alone, as illustrated in Table 2. At a creatinine concentration of 150 µmol/l obtained using the Beckman CX3 assay, and assuming the subject was a 60-year-old Caucasian male, all methods generated a GFR that placed the patient in CKD stage 3. When the creatinine concentration, obtained from the Beckman CX3 assay, fell to 120 µmol/l, for this individual the ID-MS, Roche enzymatic and compensated kinetic Jaffe assays overestimated GFR and failed to identify CKD stage 3. This was the case irrespective of whether the original or ID-MS-traceable MDRD equation was used.
Others [12,1922] have illustrated the impact of analytical differences on GFR estimations. Some authors have indirectly calibrated their assays to reflect those of the AASK laboratory [23], and some major healthcare providers have also adopted this approach [24]. This is not ideal as the Beckman Astra CX3 method clearly demonstrates significant positive bias compared with the ID-MS reference methodology [12]. One practical solution could be to adjust the equation for assay differences. For example, using the data generated in this study, MDRD-consistent GFR estimates can be derived using creatinine concentrations obtained by the Roche compensated Jaffe method by substituting the regression coefficients (Table 1) into the MDRD equation, i.e. GFR (ml/min/1.73 m2) = 186x {[(serum creatinine + 17.5)/0.9566] x 0.011312}1.154 x age0.203 x 1.212 (if black)x0.742 (if female), thus effectively adding back in the compensation factor and adjusting for calibration differences. The more logical extension to this solution would be to use the ID-MS-traceable MDRD equation and calibrate and/or compensate creatinine assays against this ID-MS gold standard method.
Two initiatives of NKDEP may improve this situation. Work is in progress towards the development of high-level commutable reference preparations for serum creatinine calibration [13]. Secondly, the ID-MS-traceable MDRD equation has been presented [9] and is now recommended for use with methods that can demonstrate ID-MS-traceable calibration [4].
We have assessed the impact of the ID-MS-traceable MDRD equation on eGFR using three creatinine methods for which it should be suitable. Estimated GFRs calculated using creatinine data generated using our own ID-MS method gave excellent agreement with those derived from the Beckman Astra CX3 assay using the original MDRD equation. This is reassuring and validates the use of the ID-MS-traceable equation in an independent cohort of patients. The Roche enzymatic assay has been calibrated against an ID-MS reference method and also gave an acceptable agreement with the eGFRs derived from the original MDRD equation. In both cases, agreement was markedly improved by using the ID-MS-traceable equation compared with the original MDRD equation, which tended to cause an overestimation of GFR when calculated with these methods.
In the Roche Integra compensated Jaffe assay, 18 µmol/l is automatically subtracted from each result to account for reaction due to non-creatinine chromogens. The manufacturers have also adjusted the calibration of the assay (slope correction) such that although the compensated assay produces lower concentrations in the normal range compared with their uncompensated assay (e.g. 37 µmol/l compared with 50 µmol/l), at higher concentrations it produces higher results (e.g. 758 µmol/l compared with 700 µmol/l) (manufacturer's data). The point of equivalence would appear to be
185 µmol/l [8]. A similar adjustment (27 µmol/l) has been made to creatinine assays on the manufacturer's Hitachi platforms. Roche Integra and Hitachi platforms are widely available: nearly 30% of laboratories registered in the United Kingdom National External Quality Assessment Scheme (UKNEQAS) currently use these methods. The Roche compensated Jaffe assay is reported by the manufacturers to be closely aligned to their enzymatic assay and to have been standardized against ID-MS and Standard Reference Material 914. On this basis, the method should be applicable for use with the ID-MS-traceable MDRD equation. Our data demonstrate that the method still produces GFR estimates that are positively biased compared with those generated by the Beckman Astra CX3 assay using the original MDRD equation, although this overestimation is almost halved when the ID-MS-traceable equation is used.
Lack of agreement between creatinine methods leads to differences in estimated GFRs calculated using the MDRD equation that may be of significance at both an individual clinical and healthcare planning strategy level [25]. Use of the ID-MS-traceable MDRD equation improves agreement in GFR estimation for methods that are so aligned, although we have demonstrated that, for the Roche compensated kinetic Jaffe assay at least, significant bias may persist. The majority of creatinine assay results generated by clinical laboratories are not zero-biased compared with ID-MS and demonstrate differing degrees of agreement with the Beckman Astra CX3 method [12]. Some [20] have argued that errors of this magnitude at this level of GFR are unlikely to be clinically significant and that universal alignment is not a practical solution. We feel that laboratories should be able to do better. The application of the National Glycohemoglobin Standardization Programme to glycated haemoglobin measurement is a good example of international alignment that has helped to deliver improvements in patient care. It is hoped that manufacturers will increasingly move towards closer alignment of their creatinine assays to ID-MS methods, as has been urged [13]. In the interim, knowledge of method comparability against either the Beckman Astra CX3 method or ID-MS creatinine methods could be used to facilitate better uniformity in the reporting of eGFR. A feasible approach could be for national external quality assessment organizations to co-ordinate the collection of such method group comparison data.
| Acknowledgments |
|---|
We are grateful to UKNEQAS for permission to use their data and to the WellChild Trust for supporting the ID-MS determinations.
Conflict of interest statement. None declared.
| References |
|---|
|
|
|---|
- Levey AS, Bosch JP, Lewis JB et al. A more accurate method to estimate glomerular filtration rate from serum creatinine: a new prediction equation. Modification of Diet in Renal Disease Study Group. Ann Intern Med 1999; 130: 461470
[Abstract/Free Full Text] - Levey AS, Greene T, Kusek JW et al. A simplified equation to predict glomerular filtration rate from serum creatinine. J Am Soc Nephrol 2000; 11: A0828
- Lamb EJ, Tomson CRV, Roderick PJ. Estimating kidney function in adults using formulae. Ann Clin Biochem 2005; 42: 321345[CrossRef][Web of Science][Medline]
- National Kidney Disease Education Program. Rationale for use and reporting of estimated GFR, 2005. Available at: http://www.nkdep.nih.gov/professionals/estimated_gfr.pdf, accessed 2 March 2006
- Clase C, Garg A, Kiberd B. Estimating the prevalence of low glomerular filtration rate requires attention to the creatinine assay calibration. J Am Soc Nephrol 2002; 13: 28122816
- Coresh J, Astor BC, McQuillan G et al. Calibration and random variation of the serum creatinine assay as critical elements of using equations to estimate glomerular filtration rate. Am J Kidney Dis 2002; 39: 920929[CrossRef][Web of Science][Medline]
- Coresh J, Eknoyan G, Levey AS. Estimating the prevalence of low glomerular filtration rate requires attention to the creatinine assay calibration. J Am Soc Nephrol 2002; 13: 28112812; author reply 28122816
[Free Full Text] - Lamb EJ, Wood J, Stowe HJ et al. Susceptibility of glomerular filtration rate estimations to variations in creatinine methodology: a study in older patients. Ann Clin Biochem 2005; 42: 1118[CrossRef][Web of Science][Medline]
- Levey AS, Coresh J, Greene T et al. Expressing the MDRD study equation for estimating GFR with IDMS traceable (gold standard) serum creatinine values. J Am Soc Nephrol 2005; 16: 69A[CrossRef]
- Carobene A, Ferrero C, Ceriotti F et al. Creatinine measurement proficiency testing: assignment of matrix-adjusted ID GC-MS target values. Clin Chem 1997; 43: 13421347
[Abstract/Free Full Text] - Lawson N, Lang T, Broughton A et al. Creatinine assays: time for action?. Ann Clin Biochem 2002; 39: 599602[CrossRef][Web of Science][Medline]
- Miller WG, Myers GL, Ashwood ER et al. Creatinine measurement: state of the art in accuracy and interlaboratory harmonization. Arch Pathol Lab Med 2005; 129: 297304[Web of Science][Medline]
- Myers GL, Miller WG, Coresh J et al. Recommendations for improving serum creatinine measurement: a report from the laboratory working group of the National Kidney Disease Education Program. Clin Chem 2006; 52: 518
[Abstract/Free Full Text] - Vickery S, Price CP, John RI et al. B-type natriuretic peptide (BNP) and amino-terminal proBNP in patients with CKD: relationship to renal function and left ventricular hypertrophy. Am J Kidney Dis 2005; 46: 610620[CrossRef][Web of Science][Medline]
- Gillingwater SD, Turner C, Swaminathan R et al. Measurement of plasma creatinine using low-injection electrospray isotope-dilution tandem mass spectrometry. In: Martin S, ed. Proc Pathol, Association of Clinical Biochemists, London: 2000; 115
- Coresh J, Toto RD, Kirk KA et al. Creatinine clearance as a measure of GFR in screenees for the African-American Study of Kidney Disease and Hypertension pilot study. Am J Kidney Dis 1998; 32: 3242[Web of Science][Medline]
- Hollis S. Analysis of method comparison studies. Ann Clin Biochem 1996; 33: 14[Medline]
- The Australasian Creatinine Consensus Working Group chronic kidney disease and automatic reporting of estimated glomerular filtration rate: a position statement. Med J Aust 2005; 183: 138141[Web of Science][Medline]
- Coresh J, Astor BC, Greene T et al. Prevalence of chronic kidney disease and decreased kidney function in the adult US population: Third National Health and Nutrition Examination Survey. Am J Kidney Dis 2003; 41: 112[Web of Science][Medline]
- Murthy K, Stevens LA, Stark PC et al. Variation in the serum creatinine assay calibration: a practical application to glomerular filtration rate estimation. Kidney Int 2005; 68: 18841887[CrossRef][Web of Science][Medline]
- Poggio ED, Wang X, Greene T et al. Performance of the modification of diet in renal disease and CockcroftGault equations in the estimation of GFR in health and in chronic kidney disease. J Am Soc Nephrol 2005; 16: 459466
[Abstract/Free Full Text] - Wuyts B, Bernard D, Van den Noortgate N et al. Reevaluation of formulas for predicting creatinine clearance in adults and children, using compensated creatinine methods. Clin Chem 2003; 49: 10111014
[Free Full Text] - Hallan S, Asberg A, Lindberg M et al. Validation of the Modification of Diet in Renal Disease formula for estimating GFR with special emphasis on calibration of the serum creatinine assay. Am J Kidney Dis 2004; 44: 8493[CrossRef][Web of Science][Medline]
- National Kidney Disease Education Program. Creatinine assay and reporting estimated GFR. Available at: http://www.nkdep.nih.gov/about/workinggroups/010603_laboratory_minutes.pdf, accessed 2 March 2006
- Van Biesen W, Vanholder R, Veys N et al. The importance of standardization of creatinine in the implementation of guidelines and recommendations for CKD: implications for CKD management programmes. Nephrol Dial Transplant 2006; 21: 7783
[Abstract/Free Full Text]
Accepted in revised form: 11. 4.06
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
H. Hobbs, P. Stevens, B. Klebe, J. Irving, R. Cooley, D. O'Donoghue, S. Green, and C. Farmer Referral patterns to renal services: what has changed in the past 4 years? Nephrol. Dial. Transplant., June 17, 2009; (2009) gfp289v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. S. Stel, F. W. Dekker, D. Ansell, H. Augustijn, F. G. Casino, F. Collart, P. Finne, G. A. Ioannidis, M. Salomone, J. P. Traynor, et al. Residual renal function at the start of dialysis and clinical outcomes Nephrol. Dial. Transplant., June 10, 2009; (2009) gfp264v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Botev, J.-P. Mallie, C. Couchoud, O. Schuck, J.-P. Fauvel, J. F.M. Wetzels, N. Lee, N. G. De Santo, and M. Cirillo Estimating Glomerular Filtration Rate: Cockcroft-Gault and Modification of Diet in Renal Disease Formulas Compared to Renal Inulin Clearance Clin. J. Am. Soc. Nephrol., May 1, 2009; 4(5): 899 - 906. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Carter, S. E. O'Riordan, G. L. Eaglestone, M. P. Delaney, and E. J. Lamb Bone mineral metabolism and its relationship to kidney disease in a residential care home population: a cross-sectional study Nephrol. Dial. Transplant., November 1, 2008; 23(11): 3554 - 3565. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Carter, S. E. O'Riordan, G. L. Eaglestone, M. P. Delaney, and E. J. Lamb Chronic kidney disease prevalence in a UK residential care home population Nephrol. Dial. Transplant., April 1, 2008; 23(4): 1257 - 1264. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. P. Quinn, A. Rainey, K. J. Cairns, A. H. Marshall, G. Savage, F. Kee, A. Peter Maxwell, E. Reaney, and D. G. Fogarty The practical implications of using standardized estimation equations in calculating the prevalence of chronic kidney disease Nephrol. Dial. Transplant., February 1, 2008; 23(2): 542 - 548. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Mariat, N. Maillard, M. Phayphet, L. Thibaudin, S. Laporte, E. Alamartine, and F. Berthoux Estimated glomerular filtration rate as an end point in kidney transplant trial: where do we stand? Nephrol. Dial. Transplant., January 1, 2008; 23(1): 33 - 38. [Full Text] [PDF] |
||||
![]() |
E. J. Lamb, M. C. Webb, and S. E O'Riordan Using the modification of diet in renal disease (MDRD) and Cockcroft and Gault equations to estimate glomerular filtration rate (GFR) in older people Age Ageing, November 1, 2007; 36(6): 689 - 692. [Full Text] [PDF] |
||||
![]() |
W. Van Biesen, D. De Bacquer, F. Verbeke, J. Delanghe, N. Lameire, and R. Vanholder The glomerular filtration rate in an apparently healthy population and its relation with cardiovascular mortality during 10 years Eur. Heart J., February 2, 2007; 28(4): 478 - 483. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||