This long-term study on the outcome of permanent silicone single-lumen dialysis catheters consisted of 43 surgically inserted catheters in 33 patients. All catheters were attached to a pressure-pressure single-cannula dialysis system. Technique and catheter survival were 80 and 59% at 1 year, and 63 and 50% at the 3rd year respectively. These parameters were not different in older patients (
70 years, n=12). Factors corresponding to a less successful outcome were placement in the left jugular vein (3-month actuarial survival of 44%) and absence of distal side holes (1-year actuarial survival of 44%). Insertion in both subclavian veins was not different to the classical right jugular vein. Infectious complications (incidence of 0.72 per 1000 catheter days) often caused simultaneous catheter and technique failure (14 versus 15%), whereas inadequate flow (incidence of 2 per 1000 catheter days) caused catheter rather than technique failure (30 versus 12%). The incidence of both complications declined significantly with time after insertion during the follow-up period. The first 3 months after insertion were more predisposed to the occurrence of inadequate flow than catheter-related infections. The general performance of these permanent single-lumen catheter devices justifies their long-term application in haemodialysis, even as a first choice of vascular access, especially in the elderly. Insertion in the left jugular vein and use of catheters without side holes should be avoided, providing a better catheter survival.
Keywords: haemodialysis; single-cannula dialysis; vascular access; permanent catheter; insertion site; side hole
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
D. G. Maki, D. M. Kluger, and C. J. Crnich The Risk of Bloodstream Infection in Adults With Different Intravascular Devices: A Systematic Review of 200 Published Prospective Studies Mayo Clin. Proc., September 1, 2006; 81(9): 1159 - 1171. [Abstract] [Full Text] [PDF] |
||||
![]() |