Skip Navigation


NDT Advance Access originally published online on September 26, 2007
Nephrology Dialysis Transplantation 2007 22(12):3533-3537; doi:10.1093/ndt/gfm452
This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
22/12/3533    most recent
gfm452v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in NDT
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (4)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Sungur, M.
Right arrow Articles by Caruso, L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sungur, M.
Right arrow Articles by Caruso, L.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The Author [2007]. Published by Oxford University Press on behalf of ERA-EDTA. All rights reserved. For Permissions, please email: journals.permissions@oxfordjournals.org



Exit of catheter lock solutions from double lumen acute haemodialysis catheters—an in vitro study*

Murat Sungur1, Emel Eryuksel2, Sinan Yavas1, Azra Bihorac1, A. Joseph Layon1,2,3 and Lawrence Caruso1,2

1Department of Anesthesiology, University of Florida College of Medicine, Gainesville, FL, USA, 2Marmara University, Istanbul, Turkey, 3Department of Surgery and 4Department of Medicine, University of Florida College of Medicine, Gainesville, FL, USA

Correspondence to: Murat Sungur, MD, Department of Anesthesiology, PO Box 100254, Gainesville, FL 32610-0254, USA. Email: msungur{at}anest.ufl.edu



  Abstract

Background. Double lumen dialysis catheters are routinely heparin or citrate ‘locked’ to maintain patency. Heparin lock-related bleeding episodes and antibiotic lock-related toxicity have been reported. The aim of this study is to quantify the amount of leak during ‘lock’ procedures and to compare leakage for different double lumen dialysis catheters.

Methods. In an experimental, in vitro study at a University research laboratory, five different double lumen dialysis catheters were tested using three different lock volumes.

Results. Using the catheter flush volume, leak ratios for Flexxicon II 15 cm and 20 cm catheters were greater than that seen in the Arrow 16 cm catheter (P < 0.05). Using 20% less than the catheter flush volume, the Flexxicon IITM 20 cm catheter had greater leak than the Duo-flow 15 and 20 cm catheters and Arrow 16 cm catheter (P < 0.05). The Flexxicon IITM 15 cm catheter had greater leak than the Duo-flow 15 cm and Duo-flow 20 cm catheters with 20% less locking volume (P < 0.05). Using 20% greater than the catheter flush volume, the Duo-flow 20 cm catheter had significantly less leak ratio than the Flexxicon IITM 20 cm catheter (P < 0.05). There were no other significant differences in leak ratios between the catheters.

Conclusion. All double lumen dialysis catheters we tested have a substantial amount of leak even when the catheter ‘lock’ volumes were used, and leak ratio increases significantly with 20% overfill. There is a leak even when using 20% less ‘lock’ volume. The amount of leak can be clinically important and may explain the reports of bleeding episodes after heparin lock and antibiotic toxicity after antibiotic and anticoagulant combination lock. Some devices have lower leak ratios than others, likely related to catheter design.

Keywords: complications; dialysis catheter; heparin; heparin leak


*A part of this study was presented in the 10th Annual Conference on Continuous Renal Replacement Therapies, Dan Diego, California, 10–12 March 2005.

Received for publication: 1. 2.07
Accepted in revised form: 12. 6.07


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?

Related articles in NDT:

In this issue ...

NDT 2007 22: i. [Extract] [FREE Full Text]  



This article has been cited by other articles:


Home page
Nephrol Dial TransplantHome page
I. Markota
Reply
Nephrol. Dial. Transplant., July 1, 2009; 24(7): 2295 - 2295.
[Full Text] [PDF]


Home page
Nephrol Dial TransplantHome page
I. Markota, D. Markota, and M. Tomic
Measuring of the heparin leakage into the circulation from central venous catheters--an in vivo study
Nephrol. Dial. Transplant., May 1, 2009; 24(5): 1550 - 1553.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.