Nephrol Dial Transplant (2003) 18: 1174-1180
© 2003 European Renal Association-European Dialysis and Transplant Association
Reducing vascular access morbidity: a comparative trial of two vascular access monitoring strategies
1 Department of Medicine, University Health Network, The Toronto General Hospital and 2 Clinical Epidemiology Unit, Sunnybrook and Women's College Health Sciences Centre, Toronto, Canada
Background. Thrombosis is the primary cause of access failure in polytetrafluoroethylene grafts and arteriovenous fistulas. It can lead to significant patient and access morbidity and mortality, and is difficult to prevent medically. Intervention is largely limited to maximizing access patency by detecting culprit lesions early and intervening with angioplasty or surgical revision. The most efficacious monitoring strategy is undetermined.
Methods. This 3 year prospective study took advantage of a change in monitoring strategy used in a large dialysis centre to compare the efficacy of two methods used to monitor grafts and fistulas in order to prevent access thrombosis. Accesses were monitored using Duplex ultrasonography in year 1, while the saline ultrasound dilution technique (Transonic) became the primary monitoring strategy in year 3 (year 2 was a transition year). Risk factors for thrombosis were determined using multivariate survival analysis, and the performance of Duplex ultrasonography and Transonic monitoring was assessed.
Results. A total of 303 656 access days at risk were assessed, with 344, 385 and 425 accesses in years 1, 2 and 3, respectively. The total thrombosis rate was 1.01/1000 access days in year 1 compared with 0.66/1000 access days in year 3. This was accomplished despite a reduction in procedure rates of 55% for angiograms, 13% for angioplasties and 31% for thrombolysis.
Conclusion. Low flow rates detected using Transonic monitoring were associated with increased thrombosis, while stenosis detected using Duplex ultrasonography was not a strong predictor of incipient thrombosis; however, these different access characteristics were compared using monitoring techniques that may be ideal in different clinical situations.
Keywords: access monitoring; fistulas; grafts; thrombosis; ultrasound dilution
Correspondence and offprint requests to: Charmaine E. Lok, MD, Department of Medicine, Division of Nephrology, The Toronto General Hospital, 11 EN-216, 200 Elizabeth Street, Toronto, Ontario, M5G 2C4 Canada. Email: charmaine.lok{at}uhn.on.ca
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
P. Ponce, A. Mateus, and L. Santos Anatomical correlation of a well-functioning access graft for haemodialysis Nephrol. Dial. Transplant., February 1, 2009; 24(2): 535 - 538. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Tessitore, V. Bedogna, A. Poli, W. Mantovani, G. Lipari, E. Baggio, G. Mansueto, and A. Lupo Adding access blood flow surveillance to clinical monitoring reduces thrombosis rates and costs, and improves fistula patency in the short term: a controlled cohort study Nephrol. Dial. Transplant., November 1, 2008; 23(11): 3578 - 3584. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Wijnen, F. M. van der Sande, J. H. M. Tordoir, J. P. Kooman, and K. M. L. Leunissen Effect of online haemodialysis vascular access flow evaluation and pre-emptive intervention on the frequency of access thrombosis NDT Plus, October 1, 2008; 1(5): 279 - 284. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Schuman, A. Ronfeld, C. Barclay, and P. Heinl Comparison of Clinical Assessment With Ultrasound Flow for Hemodialysis Access Surveillance Arch Surg, December 1, 2007; 142(12): 1129 - 1133. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. E. Lok Fistula First Initiative: Advantages and Pitfalls Clin. J. Am. Soc. Nephrol., September 1, 2007; 2(5): 1043 - 1053. [Full Text] [PDF] |
||||
![]() |
G. A. Knoll, P. S. Wells, D. Young, S. L. Perkins, R. M. Pilkey, J. J. Clinch, and M. A. Rodger Thrombophilia and the Risk for Hemodialysis Vascular Access Thrombosis J. Am. Soc. Nephrol., April 1, 2005; 16(4): 1108 - 1114. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Wiese and B. Nonnast-Daniel Colour Doppler ultrasound in dialysis access Nephrol. Dial. Transplant., August 1, 2004; 19(8): 1956 - 1963. [Full Text] [PDF] |
||||