Skip Navigation

This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (9)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Ragg, J. L.
Right arrow Articles by Anderton, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Ragg, J. L.
Right arrow Articles by Anderton, S.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Nephrol Dial Transplant (2003) 18: 955-960
© 2003 European Renal Association-European Dialysis and Transplant Association

Confidence limits of arteriovenous fistula flow rate measured by the ‘on-line’ thermodilution technique

Joe L. Ragg, John P. Treacy, Paul Snelling, Melinda Flack and Sonia Anderton

Northern Territory Clinical School, Flinders University, Royal Darwin Hospital, Northern Territory, Australia

Background. A method is presented for estimating the confidence limits (CLs), or accuracy, of the arteriovenous fistula flow rate measured at haemodialysis by the ‘on-line’ thermodilution technique.

Methods. This was by derivation of an expression to estimate what variance a set of repeated measures of flow would yield, using values pertaining to a single measure of flow. (Laws of variance were applied to the formula used to calculate flow, to account for its variables' values and measurement errors.) This enabled CLs of a single measure to be estimated.

Results. The variance estimated from a single measure was compared with that actually observed upon immediately taking a second measurement; differences in 189 pairs were not significantly different from zero (P=0.56). Applying the results demonstrated that measured flow values of 430–570 ml/min typically had associated 95% CLs that included 500 ml/min; therefore, true flow could not be said to be either side of 500 ml/min. The same was the case for 500–700 ml/min with regard to 600 ml/min. CLs widened considerably with the magnitude of flow rate, limiting the accurate measurement of higher flows and the detection of falls in flow.

Conclusion. A method to estimate CLs of flow rate measured by the thermodilution technique is presented and validated. Application demonstrates an accurate measurement of low flow, but limitations at higher flow and in detecting falls in flow. Appreciating the magnitude of such is critical to informed clinical decision making when using flow rate in an access surveillance programme.

Keywords: access flow; accuracy; haemodialysis; thermodilution; vascular access

Correspondence and offprint requests to: Dr J. L. Ragg, 26 Learmonth Street, Ballarat, Victoria 3350, Australia. Email: joe.ragg{at}mh.org.au


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.