Skip Navigation

This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow An erratum has been published
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (5)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Scanziani, R.
Right arrow Articles by Maroni, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Scanziani, R.
Right arrow Articles by Maroni, S.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Nephrol Dial Transplant (2003) 18: 2391-2398
© 2003 European Renal Association-European Dialysis and Transplant Association


Original Article

Intraperitoneal hydrostatic pressure and flow characteristics of peritoneal catheters in automated peritoneal dialysis

Renzo Scanziani1, Beatrice Dozio1, Ivano Baragetti2 and Serena Maroni1

1Renal and Dialysis Unit, Desio Hospital, Milan 2Department of Medicine, Division of Nephrology and Dialysis, Bassini Hospital, Cinisello Balsamo, Milan, Italy

Correspondence and offprint requests to: Renzo Scanziani, MD, Renal Unit, Desio Hospital, I-20033 Desio, Milan, Italy. Email: juscanz{at}tin.it

Background. In automated peritoneal dialysis (APD) one of the most important factors that influence the efficiency of the treatment is the total volume of dialysate infused per session and the dwell time. This study is aimed at examining the relationships between i.p. pressure (IPP), dialysate flow characteristics, and different dialysate fill volumes in order to optimize APD.

Methods. We studied 20 patients who received APD, with the standard fill volume (2 l, A), or individualized fill volumes based on the patient’s body surface area (2.5 l/BSA/1.73 m, B) or on body weight (40 ml/kg body weight, C). The patient’s tolerance to a given fill volume was evaluated by measuring IPP, and catheter flow characteristics were evaluated by an automated machine.

Results. IPP increased with the increase of the infused volume of dialysate (P < 0.05) and tended towards a positive relationship with the patient’s body mass index (BMI: A vs IPP: R = 0.39, P = 0.0019; B vs IPP: R = 0.66, P = 0.0012; C vs IPP R = 0.55, P = 0.009). We also found a relationship between fill volume, BMI and IPP: IPP = 1.0839 + 0.53 (ß) x BMI + 0.211 (ß) x fill volume (R = 0.65; r2 = 0.40 P < 0.01). The mean IPP with different dialysate fill volumes tended to be related to the volume of dialysate drained at the transition point (R = 0.37; P < 0.05). The pre-transition flow rate/mean IPP ratio tended towards a positive relationship with the volume of dialysate drained at the transition point (R = 0.35, P < 0.05), the transition time (R = 0.34; P < 0.05) and a negative one with the transition volume (R = –0.35, P = 0.05).

Conclusion. It is possible to customize APD, where the tidal percentage coincides with the transition point for a given catheter and a specific initial dialysate fill volume, the tolerance of which can be measured by assessing IPP.

Keywords: automated peritoneal dialysis; fill volume; intraperitoneal hydrostatic pressure


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
Nephrol Dial TransplantHome page
A. Dejardin, A. Robert, and E. Goffin
Intraperitoneal pressure in PD patients: relationship to intraperitoneal volume, body size and PD-related complications
Nephrol. Dial. Transplant., May 1, 2007; 22(5): 1437 - 1444.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.