Nephrol Dial Transplant (2002) 17: 462-466
© 2002 European Renal Association-European Dialysis and Transplant Association
Detection of microemboli in the subclavian vein of patients undergoing haemodialysis and haemodiafiltration using pulsed Doppler ultrasound
1 Department of Neurology and 2 Department of Internal Medicine, University of Münster, Münster, Germany
Background. The pathophysiology leading to pulmonary side effects during haemodialysis and haemodiafiltration is not yet fully understood. Chronic microembolization, which can be demonstrated by pulsed Doppler ultrasound, may be one cause.
Methods. The study cohort consisted of 24 long-term dialysis patients undergoing haemodialysis (n=21) and online-haemodiafiltration (n=3), respectively. The subclavian vein downstream to the venous access was investigated during different phases of the procedure using a 2-MHz pulsed ultrasound device.
Results. In all periods investigated (connection, dialysis, disconnection), numerous microembolic signals (MES) were found in the subclavian vein. The numbers of MES detected during haemodiafiltration (314709 MES per 10 min) were higher than during haemodialysis (081 MES per 10 min).
Conclusions. The composition (gaseous or solid) and origin (pump, tubing system or shunt) of the microemboli detected remains unclear. Chronic microembolization may be one cause of pulmonary complications of haemodialysis and haemodiafiltration. The detection method described in this article will help us to better understand this process and to determine what role microemboli might play in pulmonary and central nervous system disorders. It may also help to optimize the devices and techniques used.
Keywords: dialysis; microemboli; ultrasonography
Correspondence and offprint requests to: P.D. Dr Dirk W. Droste, Klinik und Poliklinik für Neurologie, Universitätsklinikum Münster, Albert-Schweitzer-Strasse 33, D-48129 Münster, Germany. Email: droste2{at}uni\|[hyphen]\|muenster.de
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
M. Barak and Y. Katz Microbubbles: Pathophysiology and Clinical Implications Chest, October 1, 2005; 128(4): 2918 - 2932. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. W. Droste, T. Beyna, B. Frye, V. Schulte, E. B. Ringelstein, and R. M. Schaefer Reduction of circulating microemboli in the subclavian vein of patients undergoing haemodialysis using pre-filled instead of dry dialysers Nephrol. Dial. Transplant., November 1, 2003; 18(11): 2377 - 2381. [Abstract] [Full Text] [PDF] |
||||