Nephrol Dial Transplant (1999) 14: 2433-2437
© 1999 European Renal Association-European Dialysis and Transplant Association
Improved bacteriological surveillance of haemodialysis fluids: a comparison between Tryptic soy agar and Reasoner's 2A media
1 Department of Internal Medicine, Albert Schweitzer Hospital, Dordrecht, 2 Regional Laboratory of Medical Microbiology (RLM), Dordrecht/Gorinchem, and 3 Department of Clinical Chemistry, Albert Schweitzer Hospital, Dordrecht, The Netherlands
Correspondence and offprint requests to: Dr G. M. Th. de Jong MD PhD, Department of Internal Medicine, Albert Schweitzer Hospital, PO Box 444, 3300 AK Dordrecht, The Netherlands.
Background. Accurate microbiological surveillance in haemodialysis centres is important as end-stage renal patients can suffer from pyrogenic reactions due to bacterial contamination of dialysis fluids. To evaluate the microbiological quality of haemodialysis fluids, special nutrient-poor culture techniques are necessary. Although the Association for the Advancement of Medical Instrumentation (AAMI) recommends Tryptic soy agar (TSA) as the standard agar, several studies have resulted in a general preference for Reasoner's 2A (R2A) agar, as it appeared to be more sensitive in demonstrating contamination of typical haemodialysis associated bacteria. In the Netherlands TSA is still used for culturing dialysate, while dialysis water is cultured on R2A. Therefore, the aims of our study were to evaluate bacterial yields of dialysis fluids on both media, and to qualify their use in routine microbiological monitoring within our haemodialysis centre.
Methods. Between April 1995 and March 1996, 229 samples of pre-treated and final purified dialysis water, and samples of dialysates were collected. The specimens were aseptically taken from the tap, various points of the reverse osmosis (RO) water-treatment system, and the effluent tubes of 32 bicarbonate haemodialysis machines. Samples of 0.1 ml were inoculated in duplicate on spread plates with TSA and R2A agars. After 10 days of incubation at 25±2°C, the numbers of colonies were quantified. The ranges of spread were taken 0100 and 0200 colony-forming units per milliliter (c.f.u./ml).
Results. The R2A agar had significantly higher colony counts than TSA agar for both dialysis water and dialysates. Considering 100 c.f.u./ml as the upper allowable bacterial limit for all dialysis fluids, microbiological non-compliance (bacterial growth) would be missed in 16% when using only TSA media (TSA
100 c.f.u./ml and R2A >100 c.f.u./ml), while this was 3% when using only R2A (TSA >100 c.f.u./ml and R2A
100 c.f.u./ml, P<0.0001). Considering 200 c.f.u./ml as the upper limit, non-compliance would have been missed in 10% when using only TSA (TSA
200 c.f.u./ml and R2A >200 c.f.u./ml), and 2% when using R2A (TSA >200 c.f.u./ml and R2A
200 c.f.u./ml, P=0.0011).
Conclusions. Microbiological surveillance of haemodialysis fluids, including pre-treated dialysis water samples collected from RO treatment systems, can be performed more precisely with R2A media than TSA, when incubated at 25±2°C for 10 days.
Keywords: bacterial contamination; culture media; haemodialysis; pyrogen; water purification
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
D. W. Griffin Atmospheric Movement of Microorganisms in Clouds of Desert Dust and Implications for Human Health Clin. Microbiol. Rev., July 1, 2007; 20(3): 459 - 477. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Bartoloni, N. B. Porteous, and L. A. Zarzabal Measuring the validity of two in-office water test kits J Am Dent Assoc, March 1, 2006; 137(3): 363 - 371. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Ledebo On-line Preparation of Solutions for Dialysis: Practical Aspects Versus Safety and Regulations J. Am. Soc. Nephrol., January 1, 2002; 13(90001): S78 - 83. [Abstract] [Full Text] [PDF] |
||||